Creatinine

METHOD OF ESTIMATION - ALKALINE PICRATE METHOD 

  • Alkaline picrate method is used to estimate the creatinine concentration in serum. But it measures some non-creatinine chromogens present in plasma. It can be rectified by using Lloyd’s reagent, which measures only the true creatinine.

Principle

  • Creatinine reacts with the picric acid to produce a colored compound, creatinine alkaline picrate. The change in absorbance is proportional to creatinine concentration.

Reagents composition

  • Creatinine dye reagent
    • Picric acid - 8.73 mmol/L
    • Surfactant
  • Creatinine base reagent
    • Sodium hydroxide - 300 mmol/L
    • Sodium phosphate - 25 mmol/Lab

Laboratory procedure

Standard

Sample

Working reagent

1000 µL

1000 µL

Standard

100 µL

-

Sample

-

100 µL

Mix and read the optical density (T1) 6 seconds after the sample or standard addition. Exactly after 6 seconds after the first reading take second reading (T2).

Caculation

Formulae                                                           

Last modified: Wednesday, 30 November 2011, 7:36 AM